The just we need is to find a marker image, which labeled the background and the holes in different value, then find the holes position set the value with object; To label the image, 1. get the implement of the image; 2. apply the region grow find the connected object, define the object which connect with edge of the image as background, label a value to express the background, others labeled a value express the hole; 3. find the hole position and take place the value with the object value.
It’s obvious easy to understand than the morphology method to fill hole.
MRI take home point (11) Done
- Most artifacts in PE direction (such as motion) but chemical shift artifact which in FE direction.
- Ringing artifacts by insufficient sample points, Wraparound artifacts by insufficient FOV.
- Gd-chelates’ primary effect is marked shortening of T1; thus heavily T1-weighted sequences are essential for high sensitivity to cancer.
- Gd-chelate contrast agents have the potential to cause a new, potentially serious disease (nephrogenic systemic fibrosis or NSF) in patients who have moderate to end-stage kidney disease at the time of administration of contrast agent
MRI take home point (10)
- Image SNR increases linearly with pixel size.
- When two similarly acquired images are added or subtracted, the resulting image has a noise level that is square root of 2 higher than that in either image.
- The larger the lesion area or contrast, and the lower the noise, the easier the lesion is to detect.
- The rose model describes the relationship between lesion area, contrast, and noise for lesion detection; the larger the lesion area or contrast, and the lower the noise, the easier the lesion is to detect.
- Artifacts can be caused by imperfection of hardware including B0, B1, gradients, shielding, RF coils.
- Cross-talking in multi-slice MRI. Partial-volume artifacts.
MRI take home point (9)
- “Fast STIR” imaging collects multiple phase-encoding views (or lines in k-space) for each
- 180-TI-90-TE/2-180-TE/2-ADC pulse train, making it a practical alternative
- to FSE for obtaining fat-suppressed T2-weighted images.
- Noise is best measured as the standard deviation in the background of an MRI.
- Image SNR increases approximately linearly with static magnetic field strength above 0.5 tesla.